Category: NRG

bEnd3 culture for BBB mimetic (open static system)

Hi all, Currently I am working on a cell culture system to mimic the blood-brain-barrier (BBB). Essentially, growing endothelial cells (bEnd3) on one side of the membrane, and pericyte and/or astrocyte on the other. Below is the layout of the…

Posted in NRG

Comparing TEER values from microfluidic device with transwell

I spent some more time in understanding the numbers I am getting for our device simulations. In my simulations, I am assigning a conductivity value for my cell layer. Technically, from this information and by knowing the thickness and area…

Posted in NRG

TEER in a microfluidic gut-on-a-chip device

Paper can be found here I am working on developing an analytical model to understand how to accurately post-process the resistance values I obtain from my TEER measurements. This paper addresses a similar issue and comes up with a correction factor…

Posted in NRG

Thermal Stress and NPN Tent Transfer

From my last two posts (Anodic Bonding, Ozone Bonding), it is clear that we have some sort of bonding going on between the freestanding membrane and the posts intended to support the membrane. However, we do not understand the nature…

Posted in NRG

Burst Pressure NPN, thermal oxide, and PBS exposure

Hemodialysis chips are not surviving the priming process and the 4 hour dialysis treatment. We want to verify that the membranes are being weakened and also to determine if coating with oxide will ameliorate the problem. Using single slot chips…

Posted in NRG

Ozone Bonding Vincent’s Chips SEM

We have wondered if we can skip anodic bonding and bond the chips using our UV ozone process. There are pros and cons to this approach, mainly that the setup and procedure is easy, but the contact and alignment is difficult.…

Posted in NRG

Variation of TEER values during cell growth in a transwell with chopstick electrodes

This is an addendum to the results I posted in my previous blog. I am trying to simulate cell growth on a transwell device and perform TEER measurements using a small chopstick electrode pair configuration. I modeled transwell membrane with…

Posted in NRG

Variation of TEER values during cell growth in microfluidic device

In my previous posts so far, I performed simulations to understand the behavior of TEER variation upon changes in electrode and membrane configuration. None of those models addresses the effects of cell growth on the TEER values. In real experiments,…

Posted in NRG

Simple Sieving Model

In an effort to better understand and check Karl’s model of sieving, I’ve built a significantly simpler version. The goal is to investigate the potential benefits of thinner membranes in dead-end filtration and separations. In my model 1 ml of…

Posted in NRG

Anodic Bonding Vincent’s Chip Stack

Previous posts: Anodic Bonding Theory, Anodic Bonding Setup   I had originally hoped there would be enough residual impurity in the oxide layer to create a direct bond with Vincent’s chips. This was a foolish hope. Pyrex is at least…

Posted in NRG