Multiple Chips packaging Concept

While my focus will be on packaging a single multi-channel chip device, I have captured here an initial concept for packaging multiple channels. The technique required is the application of verticle PDMS structures to direct blood and dialysate to and

Posted in NRG Tagged with: , ,

The effect of pump rate and channel height on Urea clearance (COMSOL modeling)

Comsol modeling was done to find the ideal channel height and volumetric flow rate. What is clear from the simulation is that ever increasing flow rates in the model lead to increased clearance and decreasing channel heights give some benefit

Posted in NRG Tagged with: , , ,

Urea clearance in the presence of BSA in PBS

In this experiment I use a syringe pump to move urea dissolved in ddH2O through a 10 mm long, 300 µm high, 500 µm wide dialysis channel. The retentate is collected in a fraction collector then measured using a urea

Posted in NRG

Adsorption of F-BSA on PEGylated pnc-Si, PEGylated Glass, Glass

This study was run to test test the efficacy of PEGylation in blocking BSA adsorption. Three surface types were used, glass, PEGylated glass, and PEGylated pnc-SI. The glass was cleaned in Ethyl Alcohol, dried and UV treated, PEGylated glass was

Posted in NRG Tagged with: , , ,

Long Term Urea Clearance (8 hours)

SET-UP Syringe pump: 5 mL of 7.6-nmol/100µL, 5.6 µL/min pump rate. Single Channel dialyser, 10 mm long, 0.5 mm wide, 0.3 mm high in stirred beaker of DI water Fraction Collector programed to rotate every 30 min. Set up in

Posted in NRG Tagged with: , ,

BSA Retention in Single Channel Dialyzer

Regarding the single channel dialyzer chip with 20-nm thick nano-porous silicon membrane: 10 mm long, 300 µm deep, 500 µm wide channels. Now that we’ve shown that a measurable quantity of urea is filtered out of DI water in one

Posted in NRG Tagged with: , ,

Single Channel Dialyzer

The main difficulty with creating a simulation of the single channel dialysis chip is that the membrane is so very thin (20 nm). This is especially true when compared to its length of 10 mm but also when compared to

Posted in NRG Tagged with: , ,

Single Channel 20 nm membrane breakage…

There is a reliability/handling issue with the single slot chips. 0.5 mm x 10 mm slots, 20 nm thick membranes with 40 nm SiN reinforcement. The membranes are easily packaged in an upper fluidic PDMS layer with inlet/outlets and two

Posted in NRG Tagged with: , ,

Rapid Prototype of Single Channel Dialysis Device

In order to rapidly test the entire setup for dialysis evaluation, including piezoelectric micro motor and recirculating fluid path, a means of quickly evaluating PDMS designs is needed. A mold was created with a multi-print technique in which printer toner

Posted in NRG Tagged with: , , ,

Urea Reduction Ratio, URR

As discussed in a previous post “Clearance, Kt/V, and Concentration…”, we will use the Kt/V metric to measure the effectiveness of dialysis. There are other methods that are used: one of these is the Urea Reduction Ratio, URR.  URR is